This manuscript presents a microfluidic Push-Pull probe for scanning electrochemical microscopy
(SECM) consisting of a working microelectrode, an integrated counter/reference electrode and two
microchannels for pushing and pul領(lǐng) an electrolyte solution to and away from a substrate. With such a
configuration, a droplet of a permanently renewed redox mediator solution is maintained just at the
probe tip to carry out SECM measurements on initially dry substrates or in microenvironments. For
SECM imaging purposes, the probe fabricated in a soft polymer material is used in a contact regime.
SECM images of various gold-on-glass samples demonstrate the proof-of-concept of a Push-Pull probe
for local surface activity characterization with high spatial resolution even on vertically oriented
substrates. Finite element computations were performed to guide the improvement of the probe sensitivity.